AutoOOF Guide#
Context#
AutoOOF is recommended for observing at frequencies of 40 GHz and higher and only available for use with Rcvr26_40 (Ka-Band), Rcvr40_52 (Q-Band), Rcvr68_92 (W-Band), RcvrArray75_115 (Argus), and Rcvr_MBA1_5 (MUSTANG-2).
In this guide, you will learn about the steps of the OOF process and how to interpret the results. For information on OOF data display in AstrID, see OOF Data Display. For an explanation of what OOF is conceptually see An Explanation of OOF.
Note
Extended sources may be observed without the AutoOOF corrections if the science is not impacted by the primary beam variations.
Finding a Good OOF Source#
What is a good source to OOF on?
You want to choose a source that is bright, compact, and in the elevation range of 30 < el < 75 (e.g., a bright calibrator). The source should be (preferably) at least 7 K in the observed band, which is about 4 Jy at Q-band (see Table 1 for the recommended flux limits for the receivers that can OOF).
Receiver |
OOF Frequency (GHz) |
Recommended Flux Limit (Jy) |
Focus Offset (mm) |
|---|---|---|---|
Ka |
32 |
3.5 |
38 |
Q |
43 |
5 |
35 |
W |
77 |
6 |
12 |
Argus |
86 |
7 |
10 |
MUSTANG-2 |
90 |
1 |
10 |
Note
The frequency which an OOF is taken at is hard coded. Table 1 shows the frequencies for each receiver. MUSTANG-2 is a continuum instrument and the fiducial band center of 90 GHz is used.
How do I find a good OOF source?
You can use the default CLEO catalogs to find potential bright calibrators. However, you should NOT rely on the catalog flux to be accurate as it is often many years out of date. It is preferable to instead use the ALMA Calibrator Source Catalogue to find good OOF sources as these sources are well understood and observed often. The ALMA Calibrator Source Catalogue has an extensive record of the flux densities for many of the bright 3 mm sources.
If you unsure about the brightness of your source then you can run a point/focus scan on the calibrator first in order to confirm its strength. Remember, to obtain good OOF results, you need to be able to detect the source when the subreflector is out of focus which reduces its peak intensity significantly.
Note
In the past it was true that you would want to OOF at an elevation that is similar to your source. However that was before the new gravity model implemented in 2010. Now the gravity model is good enough that this is not necessary.
Warning
Do not OOF on a source that is in the keyhole (> 85°) or about to enter the keyhole. Find a source to OOF on that has an elevation of 30-75°. Remember that an OOF can take up to 30 minutes to complete so make sure that your OOF source will not go into the keyhole during the duration of the OOF process.
Receiver-Specific OOF Guidance#
Since the Ka-band receiver with the default CCB backend provides the most accurate measurements of the surface parameters, users should consider using this whenever possible (except for MUSTANG-2 users who use the MUSTANG system for OOFing). The benefit of using the Ka+CCB system when possible is two-fold: (a) due to the high sensitivity provided by the CCB, the S/N ratios observed with the Ka+CCB system are much higher than what is possible at higher frequencies, and (b) the winds affect the Ka-band data to a lesser degree due to the larger beam-size which in turn makes the surface solutions less affected by winds.
Q-Band#
In general the advice is to use Ka when you can, but if you are observing with Q-band, just use Q. You don’t gain that much by switching to Ka.
W-Band#
See guidance in W-band Observing, but in summary if the Ka-band receiver is available, run AutoOOF() at Ka-band (or if Ka is not available and Q-band is available, use Q-band) instead of W-band for more accurate surface corrections.
Argus#
See guidance in Argus Observing, but in summary, again, if the Ka-band receiver is available, run the AutoOOF at Ka-band instead of Argus for more accurate surface corrections. Then after the AutoOOF solutions are applied, run a point and focus with Argus to confirm the telescope collimation offsets after the application of the OOF solutions.
MUSTANG-2#
Use MUSTANG-2 to OOF and obtain pointing and focus corrections. For choosing your OOF sources for MUSTANG-2,
see guide on picking OOF sources.
In the AutoOOF(), if it is the first OOF for the MUSTANG-2 project set
calseq=True so AutoOOF(source,calseq=True). When you set calseq=True, this initiates a skydip
scan before the OOF scans (skydip is needed for calibration of the data). If you are doing an OOF later in
the night then set calseq=False so that you do not run another skydip.
The AutoOOF Procedure#
When you come online, check if there were high frequency observations before you at any point that did an AutoOOF. You can ask the operator about this. If there were no high frequency observations before you, the Zernike thermal coefficients should be zeroed out and the Zernike thermal offsets turned off. The operator should do this before the first high frequency project of the night, but it is good practice to check this. Additionally, if the OOF was done before ~22:00, you may want to request the same thing (that the Zernike thermal coefficients are zeroed out and the Zernike thermal offsets are turned off) before you OOF, because the dish may still be settling and the previous corrections may not be similar to what you calculate.
For a more in-depth explanation of what OOF is doing see this section of the OOF explanation.
Below we provide some general information about the OOF process:
-
- Allow approximately 25 minutes for an AutoOOF.
-
The AutoOOF procedure obtains three consecutive OTF maps; each map takes 5-6 minutes and is at a different focus position.
-
- Use AutoOOF to derive pointing and focus offsets
-
The OOF-processing is launched automatically upon completion of the third map, and the result is displayed in the OOF tab of AstrID. It is incumbent upon you, as the observer, to examine the solutions (see guidance in guide below), select a solution, and click the button (in the AstrID DataDisplay tab) to send the selected solution to the active surface. It is recommended that when sending the solutions, you use the button in the OOF display tab labeled After selecting the Zernike solution above (e.g., z4 or z5), click this green button to send the solutions to the telescope.
Note
AutoPeakFocus() may be run as a sanity check on the AutoOOF solution. If Peak/Focus scans were performed before AutoOOF, then source amplitude should be greater after the AutoOOF than what was seen before the surface correction was sent. Additionally, AutoPeakFocus() pointing and focus corrections should agree with values derived by AutoOOF().
How long does an OOF solution remain valid?#
AutoOOF() is currently only executed for observing projects at nighttime.
Nighttime#
A general rule of thumb is that if the corrections are measured at least two hours after sunset, then the solutions should be good for about 4 hours. This depends on how rapidly the backup structure cools off after sunset and how sunny the day was. If the OOF is taken after midnight, the structure has typically stabilized by then, and the solutions may even be good until after sunrise. But in general, OOF solutions are good for 2-6 hours at nighttime depending on the conditions of any particular night. Note that if you are scheduled soon after sunset, be aware that you might have to OOF more often as the surface “settles.”
Daytime#
During the daytime, it is a difficult to answer the question of “How long does the OOF solution remain valid?” as it depends on the position of the telescope with respect to the Sun and cloud cover. The answer can be anything from less than 30 min to 4 hours. Daytime surface changes are on time scales of the order of <1 hour. Due to these rapidly changing conditions, the AutoOOF() solutions (which are dervied on similar timescale of ~30 minutes) can cause more harm (efficiency damage) than good. So it is typically not useful to use the OOF derived corrections during the day.
When do I need to OOF again?#
No matter what receiver you are using you should periodically examine peak scans.
- A new AutoOOF may be necessary if the following characteristics are seen:
-
Significant sidelobes begin to appear.
The beam size increases by more than 10%.
Source amplitude decreases systematically by 15% or more.
For how often to do a peak scan for Ka-Argus receivers see See Table 1 in Pointing and Focus Strategies. MUSTANG-2 observers should observe a pointing calibrator and use it to check the peak and beam approximately every 30 minutes.
Todo
Using the Ruze and radiometer equations, calculate at what point the surface degrades enough that it will be more efficient use of time to OOF as opposed to keep observing
Inspecting AutoOOF Solutions#
Surface Delta Map Characteristics#
- Good solutions have the following characteristics:
-
Broad features of less than \(\pm 1.5\) radians of phase (the amplitude of the features is in radians and indicated by the colorbar on a surface delta map - see Fig. 3) in good, stable conditions. Note that you may uncheck Show Fixed Scale Image to view the full data range in the color bar.
Surface RMS \(\lesssim\) 400\(\mu m\) if the gravity-Zernike model is on and the temperature is greater than 10° F.
- If you see an RMS \(\gtrsim\) 400\(\mu m\), the solutions might still be ok to apply. The following are conditions in which you would expect to have a higher RMS:
-
The temperature is < 10° F. The gravity-Zernike model is an average model taken in a variety of conditions (see Modeling and Correcting for Gravity). Consequently, if the conditions under which you are performing the AutoOOF procedure differ substantially from “typical” conditions (defined by the gravity-Zernike model), the resulting solutions exhibit corresponding deviations. For example, the gravity-Zernike model was taken in the temperature range of 20-40° F. Therefore, at temperatures < 10° F, you have begun to deviate from the typical range of the gravity-Zernike model and you will see larger deltas and therefore higher RMS values.
The previous OOF solution is no longer applicable. This can happen if the OOF before you was done before ~22:00. In this case the dish may have still been settling, but is now stable when you are observing. And in this case, the differences between the previous surface and yours in stable conditions might be quite large.
The gravity-Zernike model is not on. You can check this by doing the following: open the CLEO Active Surface application and in the Standard tab (default when the Active Surface application opens) the gravity-Zernike model is on if the box next to Zernike Offsets is green (see more details in OOF Coefficients in Active Surface Application).
Here is an example of an acceptable OOF solution:
Fig. 3 This solution shows broad features (\(\pm\) 1.5 radians of phase) with a surface RMS of 197\(\mu m\).#
Here is an example of an unacceptable OOF solution:
Fig. 4 This solution shows steep contour lines (\(\pm\) 15 radians of phase) and a surface RMS of 626\(\mu m\). This is likely the result of poor quality raw data and the solution should not be used.#
Note
The scalebar in the surface delta map has units of radians. The units of the x and y axis of the surface delta map are meters.
Selecting the Zernike order#
By default, AutoOOF will halt processing after the fifth-order Zernike (z5) solution has been computed. The z5 solution is suitable for most conditions and is generally what observers should expect to use. A more aggressive sixth-order (z6) fit may also be derived at the cost of a few additional minutes of processing time. This is usually unnecessary and should only be done on bright calibrators under favorable weather conditions. See OOF z6 Processing Options for information on how to do this.
Occasionally, it may be necessary to drop to a lower order of fit if the following features are seen:
Large deviations over a significant area of the dish edge in the OOF solution.
Regularly spaced features around the circumference of the dish at higher order fits in the OOF solution.
-
Anomalous values in the pointing/focus LPC/LFC for one particular solution, or a significant jump in LPC above a certain Zernike fit order. For example, if the focus (LFCy) values for the z3-z4 solutions are around -3mm, then abruptly jump to +10mm for the z5 solution, then it would be prudent to assume that some or all of the solutions may be invalid. It may be possible to determine which solutions are valid by examining the fitted beam maps for obvious artifacts or deviations from the observed beams.
Fig. 5 The AutoOOF fitted beam maps. The observed beams are plotted on the top row with the z3, z4 and z5 fits to the observed beams plotted below. The z3 solution (\(2^{\text{nd}}\) row down) shows an obvious artifact and should not be used.#
Fig. 6 Zernike Solutions. Note the significant jump in LPC and the LFC between the z3 and z4 solutions.#
Todo
remake Fig. 6.
OOF z6 Processing Options#
Deriving the sixth-order Zernike (z6) solution will require a few additional minutes of processing time. To do this, the user will have to manually change the maximum order of fit to process to z6 in the following way:
Select the OOF Subtab of the AstrID DataDisplay.
Select Tools \(\rightarrow\) Options… from the drop–down menu.
-
Select the maximum order of fit to process from the Processing Options tab of the pop–up window.
Important
All changes must be made before submitting the AstrID script containing the AutoOOF() function in order to take effect. You may also repeat processing after making any changes by pressing Reanalyze OOF (Online Only.
Internal Access Only
More information on AutoOOF can be found here.
Todo
Transfer the relevant content from that wiki page here.
AutoOOF Raw Data#
Although an OOF solution may appear to be reasonable (e.g., Fig. 3) it may be invalid if it was derived from a bad set of raw data. Sending such a solution to the active surface could degrade performance. Therefore, you should always check the quality of the raw AutoOOF data in order to determine whether your derived solutions are valid.
- For a set of raw data to be considered valid, it should show the following characteristics:
-
Detection of the source in the raw data timestream at all focus positions and at least two focus positions have good signal-to-noise.
Symmetrical positive/negative pattern in all three raw data images.
Smooth features in all three raw data images. Sharp edges or stripes indicate hardware/software glitches or excessive winds.
You can view the AutoOOF raw data by selecting the raw data button in the upper-right section of the OOF Subtab of the Data Display. Each column represents one focus position. The top row is the raw timestream data from the receiver, the second row has the baselines removed, and the bottom row shows the corresponding beam maps. See Fig. 7 and Fig. 8 for a comparison of acceptable and unacceptable raw AutoOOF data.
Fig. 7 A plot of the raw OOF data on a fairly clean Ka-Band+CCB dataset.#
Fig. 8 A plot of the raw OOF data on a source which is too faint in particular in the delta focus = -38 mm and = +38 mm.#
Raw Data: What am I looking at?#
For all receivers besides MUSTANG-2, each OOF is an RALongMap so each of the spikes in the raw data is the beam going over the source; and the relative strength of the spike is how much of the source is in the beam. The closer you are to the correct focus, the less fuzzy and wide the source is, and thus then the less spikes there are. Said another way, when you are close to the focus correction values, there are fewer, but stronger peaks. Further, where you are getting the spikes in time tells you about the elevation offset. For example if you see spikes early in time that means there will be a minus elevation offset.
Note that if a receiver has 2 or more beams, two beams are used to OOF. In the raw data plots you will see the signal of the source twice: once for each beam. One beam is subtracted from another to make the raw data plots so that is why there are negative spikes. And if you zoom in you will see that the positive and negative spikes aren’t exactly lined up. It is because the two beams are going over the source at slightly different times.
How do I know if I have a good SNR for the OOF? Ask yourself, do you see peaks? If yes, you have high SNR. If you just see noise, you don’t have high SNR. Typically the SNR will be highest in the OOF scan that is closest to in focus (look at what the focus solution is, the LFCy, and whichever delta focus column is closest to the focus value).
Tip for determining if you have enough SNR for OOFing.
Change to the DataDisplay -> Continuum tab for the first OOF scan and see if you see signal from a source or not. If you don’t see spikes, then you might want to abort and change to a stronger source.
Note
The y-axis in the raw data plots are NOT scaled between the three focus positions.
AutoOOF Fitted Beam Map#
It is good to quickly check the fitted beam map.
Fig. 9 A plot of the fitted beam map for an OOF taken with Argus.#
The top row shows the observed beams at the various focus positions. Each subsequent row presents the fitted beam maps for the 3rd to 5th Zernike orders if these corrections were applied. In general, higher Zernike orders yield a more accurate beam solution. Moreover, the beam appearance should become more circular near the correct focus; for example, when LFC_Y is −9 mm, the −10 mm fitted beam maps should appear more circular. Note, however, that these maps depict how point sources would look at each focus position, so the maps corresponding to two of the three focus positions are likely to deviate from the actual focus position and may appear non-circular.
Note that when the telescope is out of focus, you are stretching the beam out. Thus, being at different focus positions will stretch the beam out in different ways. Let’s say that 0 mm is at focus, then the plus focus position stretches the beam out vertically then the minus focus stretches out the beam horizontally. For example, see the fitted beam map in the bad low SNR example.
Note
There are always two beams in the fitted beam maps for all receivers that you can OOF with besides MUSTANG-2.
Summary for checking your OOF solutions#
Go through the various Zernike order fits (z5-z3) and check the following:
-
- surface delta map. If the structure of surface delta map has:
-
smooth features = good
sharp features = bad
-
- surface RMS.
-
RMS \(\lesssim\) 400 um = good
RMS \(\gtrsim\) 400 um = potentially bad. In this case, the observer should do some sleuthing into reasons why it might be higher. See cases in which you would expect to see a higher RMS in Surface Delta Map Characteristics.
-
- local point corrections (LPCs) & local focus correction (LFC)
-
smaller numbers = good
large numbers (LFC > 10mm) = bad
Also check the:
-
- the raw data
-
high SNR = good
low SNR = bad
-
- the fitted beam maps
-
streaks and/or very non-gaussian = bad
Note
A general philosophy is to apply z5 unless you can find a strong reason as to why you should not apply z5.
Tips and Tricks#
Uncheck Fixed Scale#
You’ll notice that the z-axis values of the surface delta map are by default from -2.5 to 2.5 radians (see scale bar to the right of the surface delta map). This is because the box to the right of the surface image map labeled Show Fixed-Scale Image is checked. It can be useful to see the full range of z-axis values. To do this uncheck the box next to Show Fixed-Scale Image and it will then display the full range of values. This can be helpful in the case that the OOF is clearly bad or you are uncertain.
Scaling for a Bad OOF#
Example 1#
For more details about this OOF see the Q-Band Bad Example.
Example 2#
For more details about this OOF see the MUSTANG2 Bad Example 2.
Scaling for an Uncertain OOF#
For more details about this OOF see the Q-Band Uncertain Example.
Scaling for a Good OOF#
For more details about this OOF see the MUSTANG2 Good Example 1.
Show Solutions with Focus Removed#
When you are inspecting the surface delta map, there is an option on the right hand side to Show Solutions with Focus Removed. OOF solves for the pointing and focus corrections in addition to the surface corrections. Show Solutions with Focus Removed allows you to look at the surface corrections with the focus removed. This is particularly useful to do when you have a large focus correction (\(|\) LFC_Y \(| \gtrsim\) 10 mm) and/or the focus correction is outside of the range of the focus offsets which is receiver specific (see the “Focus Offset” column of Table 1). It is also useful to check when you are unsure if you have good surface corrections (typically when you have indications that the OOF is bad or marginal) as this allows you check the surface delta map with the focus removed to see if the underlying surface corrections are good. See the examples below.
Example 1#
See the following surface delta maps of a project that OOFed with Ka+CCB for different orders of Zernike polynomials.
You will notice that all of these surface delta map look ok (no sharp edges really), but the surface RMS for z5 is a bit on the high side. The main thing of concern is that the focus offsets are quite high (\(\gtrsim|10mm|\)). This is a good case in which it is good to inspect the surface corrections with the focus removed. To do this, check the box that says Show Solutions with Focus Removed. Once you have done that you can inspect all three orders of Zernike surface delta maps. For this case they look like the following:
Note
Show Solutions with Focus Removed shows three different surface delta maps: relative solution, absolute solution, and absolute solution with focus removed. Note that the relative solution is the one that most looks like the surface delta map that is initially displayed. The absolute solution with focus removed is the map that you want to look at.
For this example, in the main z5 surface delta map you see (a) a spherically symmetric shape that is indicative of being out of focus, and (b) a higher surface RMS. Then when you look at the solution with the focus removed (absolute solution with focus removed) there is a decent surface with the spherically symmetric shape gone and the RMS is more reasonable. This example demonstrates the use of looking at the surface corrections with the focus removed. And in this case, we would say that the z5 solutions are good (we saw that the surface solutions look good via removing the focus) and would apply those.
Note
If the focus corrections are small, you wont see much of a difference in the focus removed solutions. It is when the focus corrections are larger when the focus removed solutions will be different from the solutions. As a rule of thumb, look at the focus removed solutions when the focus corrections are around what the +-focus positions are for the receiver being used (e.g., 10mm for MUSTANG-2).
Example 2#
See the following surface delta maps of a project that did an OOF with Argus.
You will notice that all of these surface delta maps look ok (wrt to their structure), but there is the characteristic spherical feature in the surface delta map, indicative of the focus being a large contribution to the surface corrections (instead of just the surface). The focus removed solutions look like the following:
Notice that the spherical structure is now gone from the absolute solution with focus removed in all cases and we see a typical shape of the surface delta maps. In this case, the advice would be to apply z5.
Examples#
To see many examples OOF results, please see OOF Examples.
Sending a Solution to the Active Surface#
When you are ready to accept the solution being displayed it will need to be manually sent to the active surface. It is recommended that when sending the solutions, you use the button labeled Send Selected Solution with Point and Focus Corrections. If you use this option, you do not have to perform a Peak or Focus after an AutoOOF. It is still good practice to run a Peak and Focus at the beginning of your observing session unless you are using MUSTANG-2. Subsequent pointing and focus corrections may be computed via AutoOOF.
Many high frequency observers will perform Peak scans immediately following an AutoOOF to verify the surface solution. If the solution is satisfactory the LPCs and LFC from Peak/Focus scans should agree with values from the OOF solution, there should be no significant sidelobes visible in the peak scans, and Peak scans should also yield the expected beam FWHM.
FAQ#
I have a bad OOF. What should I do?#
In general, if you have a bad OOF, ask operator to zero out the Zernike thermal coefficients and turn off the Zernike thermal offsets. Once that has been done, reOOF and see if things improve.
I have a low RMS. What should I do?#
Nothing! It is not bad to have a low RMS. It just means it is a stable night and the difference in the surface between when the surface was last set and now is small. Technically, there isn’t a lower limit for OOF surface RMS solutions; though GBO staff haven’t seen one below 40 \(\mu\mathrm{m}\). See below for an example of a “low” RMS of 66 \(\mu\mathrm{m}\) from an Argus observing run:
What if OOF fails?#
Sometimes OOF may time out and you will get a red screen like this.
If you get this red screen, this means that for some reason the OOF processing failed. This type of red screen has only been seen during MUSTANG-2 observations so far. OOF processing for M2 can take a while. GFM checks the directory where the processing script puts the solutions for the files and if GFM doesn’t find the solutions after a certain amount of time, it puts up that red screen.
If you get this, start another OOF and diagnose the underlying problem.
How do I view the results of a previous OOF?#
In AstrID, select the DataDisplay tab, then go to File -> Open. Load the project folder of interest via the Location field, by entering the path and hit enter. Select the ScanLog.fits file, and press Open. As you wait for the scan details to load, go to DataDislay -> OOF. Once the scan information has loaded, find the set of OOF scans (3 RALongMaps together), and select the third scan and you should see the OOF results.